Description

Configured for high-resolution temperature measurement in machinery monitoring networks, the Bently Nevada 133835-02 (133835-02 Thermocouple Isolated I/O Module) provides direct physical/electrical execution. The hardware processing interface conditions thermal electromotive force signals derived from external thermocouple sensors, executing high-voltage galvanic separation across multi-channel inputs before routing digitised temperature variables to the system processor bus.

Hardware Specifications

ParameterSpecification
Model133835-02
BrandBently Nevada
OriginUSA
Weight0.46 kg
Dimensions241.3 mm x 24.4 mm x 163.1 mm
Operating Temp-20 to +60 deg C
Power Consumption5 W max
Input Channels4 isolated channels
Input TypeThermocouple (TC)
Temperature Range-270 to +1372 deg C
Accuracy±0.1% of full scale
Isolation Voltage1500 VDC (channel-to-channel and channel-to-ground)
Power Supply24 VDC nominal (18 to 32 VDC range)
Storage Temperature-40 to +85 deg C
Humidity0% to 95% RH (non-condensing)
Resolution16-bit
Sampling RateUp to 10 samples per second per channel
Input Impedance>10 MOhm
CMRR120 dB
Linearity±0.01% of full scale
Communication InterfaceModbus TCP/IP, RS-485

Machinery Monitoring & TSI Characteristics

The module interfaces directly with static casing and bearing thermal points to monitor long-term rotor dynamics and heat dissipation behaviors. While processing high-accuracy thermal inputs, the system internal plane maps calibration parameters alongside eddy-current probe scaling configurations and automated gap voltage validation (-10 VDC targets) benchmarks. The physical architecture leverages integrated hardware trace separation to enforce strict cross-talk suppression metrics, ensuring high-frequency adjacent loop noise does not skew the internal 16-bit analog-to-digital conversion arrays.

Frequently Asked Questions

Q: Does the 133835-02 module support live hot-swap installation inside an active monitoring chassis?

A: No. Prior to removing or mounting the 133835-02 hardware interface, technicians must isolate the 24 VDC external power bus supply. Unrestricted live insertion risks inducing thermal component shock and transient backplane data degradation.

Q: How does the 133835-02 framework respond if an external thermocouple sensor suffers a physical open-circuit failure?

A: The input channel registers the high-impedance state and triggers an upscale break detection flag, which forces the related Modbus register to a designated fault value and bypasses standard machinery protection matrices.

Field Installation Guidelines

  • Align the 133835-02 card edge with the designated chassis terminal slot guides and insert firmly until the rear locking connectors seat into the backplane matrix.
  • Route all thermocouple extension lines through dedicated metal wireways to maintain cross-talk suppression baselines and separate sensor loops from dynamic power conductors.
  • Connect the overall cable shield drain to the single-point instrument ground termination bracket while maintaining the sensor junction isolation boundary intact.

Additional information

Weight0.46 kg
Dimensions85 × 90 × 70 mm

0.0 Average Rating Rated (0 Reviews)

Rated
0%
Rated
0%
Rated
0%
Rated
0%
Rated
0%

Be the first to review “Bently Nevada 133835-02 Thermocouple Isolated I/O Module”

Your email address will not be published. Required fields are marked *

You have to login to add images.

Recently Viewed Items